Section 6: Serial I/O
12/09 15 0596-0315-01 Rev B
Section 5 Calibrating the Indicator
5.1 Introduction
Calibration can be performed entirely from the front panel or locked out with security levels 2-5 or
internal jumper JP1.
Calibration can be performed with a single calibration point, or using up to four calibration points.
SL0 SL0 SL1 SL1
SL2 SL2 SL3 SL3 SL4 SL4 SL5 SL5
JP1 JP1 JP1 JP1 JP1 JP1 JP1 JP1 JP1 JP1 JP1 JP1
MENU Out In Out In Out In Out In Out In Out In
SEtUP
Y Y Y Y Y Y Y Y N N N N
COnFIG
Y N Y N Y N N N N N N N
APP
Y Y Y Y Y Y Y Y N N N N
CAL
Y N Y N N N N N N N N N
Menu access
based on security level (SL) and internal calibration jumper (JP1)
The following flow charts show step by step instructions for a single point calibration, and a multipoint
calibration.