Section 6: Serial I/O
12/09 13 0596-0315-01 Rev B
Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the selection.
Available selections are:
OFF Tare is disabled.
ON Tare is active.
On-CLr Tare automatically clears when Gross weight returns to Zero
Hl: Overload limit: The actual value at which the display goes to ol (represented as a percentage of capacity).
The display will indicate hl followed by the current setting. Press Over/Under/Next or units to scroll up, or
tare to scroll down through the available choices. Press Program/Enter to enter the selection. Available
selections are:
102.5
103 105 110 300
Ul: Underload limit: The actual value at which the display goes to ul (represented as a percentage of capacity
behind zero). The display will indicate ul followed by the current setting. Press Over/Under/Next or units to
scroll up, or tare to scroll down through the available choices. Press Program/Enter to enter the selection.
Available selections are:
5.5 10 25
50 100
D rate: Programming the display update rate: The times between display updates in seconds. The display will
indicate d rate and then the current setting. Press Over/Under/Next or units to scroll up, or tare to scroll
down through the available choices. Press Program/Enter to enter the selection. Available selections are:
0.1 0.2 0.3
0.4 0.5 0.6 0.7 0.8 0.9 1.0
P1: Output format of Port1, GTN or net only: This determines whether the output from Port1 will be
transmitted as a net weight only or a gross/tare/net weighment. The display will indicate p1 followed by the
current setting. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter
the selection. Available choices are:
Gtn nEt
P2: Output format of Port2, GTN or net only: This determines whether the output from Port2 will be
transmitted as a net weight only or a gross/tare/net weighment. The display will indicate p1 followed by the
current setting. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter
the selection. Available choices are:
Gtn nEt
The Enviro™ Series indicator will now return to the config prompt. Navigate to the desired menu heading to
continue programming, or to done to finish.
APP Menu
1. Press and hold the Program/Enter key or press S1 on the main PCB to display SEtUP.
2. Press Units or Over/Under/Next twice to display APP.
3. Press the Program/Enter key to enter APP menu.