Document Number: 0596-0320-01 Revision A
Operations setup:
Check weighing setup:
1. Press and hold the Over/Under/Next key until “CHEC” is displayed.
2. Press Enter to enter into the checkweigh setup menu.
a. Rec 1 will display
3. Press enter to accept recipe 1 or press over/under/next key to go to next recipe and enter.
4. Display shows EnA. Press over/under/next key to toggle between Y & N to enable this
recipe. Press enter to accept.
5. diS displays. Press over/under/next key to toggle between Y & N to enable the display
leds for this recipe. Press enter to accept.
6. Units will display. Press over/under/next key to toggle between the different unit
selections. Press enter to accept.
NOTE: Only those unit selections enabled in the Units setting under the “Config” menu
will be available here.
7. hAnd will be displayed. This is where you will set up the limits for the bar graph led
display for under/accept/over. Use the over/under/next key to toggle between Y and N
and enter to accept selection.
NOTE: Regardless of which selection you make, you will need to enter the values for
the end points of the three categories, under, accept, over.
a. If you select Y.
i. The leds will have the first led on the Under section lit. You will see a
value on the display. Change this value to be the low end of the Under
range of leds. Enter to accept.
ii. The leds will now have the entire under section and the first led in the
Accept section lit. The value on the display should be changed to the low
end value of the accept range. Enter to accept.
iii. The leds will now have the entire under and accept sections lit and the
first led in the over section lit. The value on the display should be
changed to the low end value of the over range amount. Enter to accept.
iv. The leds will now all be lit. The value on the display should be changed
to the high end value of the over range amount. Enter to accept.
b. If you select N
i. The leds will be scrolling back and forth. Place an acceptable weight on
the scale. This will give you a starting weight display. Press enter. The
leds for under will all be lit and the first of the accept will be lit. Adjust
the value displayed to be the low acceptable value. Press enter to accept.
ii. Only the first led in the under range will be lit. Adjust the value
displayed to be equal to the low value for the under range. Press enter to
accept.